PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 162
Using a 1/2 in. wrench, loosen the two DH200 high pressure connections on the
pressure tube between the vent valve and the thermal pressure control unit (TPCU)
Step 6
Slide back the valve assembly in the mounting slots to fully remove the
loosened tube.
Install a DH200 gland and plug (supplied with PPCH accessories) in the open
port of the TPCU.
Reapply drive air supply, connect electrical power and power on the PPCH.
Perform a pressure decay based leak check (see Section 5.11.1).
If the leak is no longer present then the high pressure vent valve is the cause
of the leak and the valve should be replaced (see Section 5.12.1).
If the leak is still present, there is a leak that is not in the vent valve.
Reconnect the original vent valve and proceed with additional leak testing to
localize a leak within PPCH (see Section 5.11.2).