Dentsply Sirona
Operating Instructions DAC Universal
5 Cleaning, maintenance and validation
5.2 Biennial Service
6681006 D3708
D3708.201.01.01.02 04.2019
105
NitraClean cleaning tablets
NitraClean cleaning tablets are used for cleaning the empty chamber
and internal cleaning of the hoses.
The empty chamber and the hoses must be cleaned when needed, or at
least in the course of Check & Clean.
IMPORTANT
NitraClean cleaning tablets must not be put in the water tank.
Please see the MSDS for further information for the NitraClean tablets.
http://www.dentsplysirona.com/en/service/safety-data-sheets/
5.2
Biennial Service
Frequency: Every 2 years or every 3000 cycles (whichever comes first).
After 2 years or 3000 cycles after the last service or initial
commissioning of the DAC Universal the message "
Service necessary"
appears on the display and, at the same time, a 2-second acoustic
signal sounds.
When this message appears, a qualified technician must perform a
service on the DAC Universal to ensure the proper and efficient
performance of the DAC Universal.