3 Product description
3.2 Controls and functional elements of the DAC UNIVERSAL
6681006 D3708
D3708.201.01.01.02 04.2019
23
3.2.3
Lid
The following lids can be used for reprocessing in the DAC Universal:
● Blue lid for reprocessing turbines, straight and contra-angle
handpieces.
See "Blue lid standard program [→ 20]" for functions and see
"Reprocessing with the Blue lid [→ 70]"
for instructions.
The maximum load on the Blue lid is 6 instruments.
Blue lid
Water consumption for cleaning
approx. 650 ml
Water consumption for disinfection
approx. 150 ml
Holding time
30 s
Holding temperature
134 °C
A0-value
>3000
● Green lid for reprocessing Green lid instruments.
See "Green lid program [→ 20]" for functions and see "Reprocessing
with the Green lid [→ 73]"
for instructions.
The maximum load on the Green lid is 6 instruments.
Green lid
Water consumption for cleaning
approx. 500 ml
Water consumption for disinfection
approx. 150 ml
Holding time
30 s
Holding temperature
134 °C
A0-value
>3000