3 Product description
3.1 Functions
Dentsply Sirona
Operating Instructions DAC Universal
20
6681006 D3708
D3708.201.01.01.02 04.2019
3.1.1
Blue lid standard program
The turbines, straight and contra-angle handpieces are reprocessed as
follows using the adapters on the Blue lid:
1.
Preliminary cleaning
2.
Leak test
3.
Internal cleaning: The internal spray and drive channels are cleaned
with water.
4.
Lubrication: The drive channels are lubricated (sufficient for the next
treatment)
5.
External cleaning: Done using the pulse-wash method (multicyclic
cleaning method)
6.
Heating up to 134°C
7.
Back-flush: Saturated steam is directed through the instruments
8.
Disinfection: 0.5 min. at 134°C
9.
Cooling
10. The lid opens slightly
11. The lid can now be open fully
3.1.2
Green lid program
The Green lid instruments are reprocessed as follows using the
appropriate adapters on the Green lid:
1.
Preliminary cleaning
2.
Leak test
3.
Internal cleaning: The internal spray channels are rinsed with water
4.
External cleaning: Done using the pulse-wash method (multicyclic
cleaning method)
5.
Heating up to 134°C
6.
Back-flush: Saturated steam is directed through the instruments
7.
Disinfection: 0.5 min. at 134°C
8.
Cooling
9.
The lid opens slightly
10. The lid can now be open fully