4 Operation
4.2 Reprocessing with the DAC UNIVERSAL
Dentsply Sirona
Operating Instructions DAC Universal
80
6681006 D3708
D3708.201.01.01.02 04.2019
Output information of the Green lid:
The following information is output via the printer:
● Date (YYYY-MM-DD)
● Time (24h) (HH:MM:SS)
● Serial number of the DAC Universal
● Process Software
● C&C 50 status (Due / OK)
● C&C 200 status (Due / OK)
● Service status
● Cycle
● Lid type
● Lid serial
● Program
For each phase the time [Time:] and the result [Res:] "OK" or "FAIL" are
printed. The phases are:
● Lid detected
● External flush
● Leakage test
● Internal cleaning
● External cleaning
● External rinse
● Heating
● Back-flush
● Start holding time
● Maximum: °C / bar
● Minimum: °C / bar
● Duration (mm-ss)
The rest of the phases and the result “OK” or “FAIL” are printed. The
phases are:
● End holding time
● Cooling
● Process end
● Confirmation
● Cycle end
The rest of the phases and the result “OK” or “FAIL” are printed. The
phases are:
● Disinfection
● Error code
If "FAIL" is displayed for a phase, then the reprocessing cycle was
unsuccessful and must be repeated.
Содержание DAC Universal Touch
Страница 1: ...New as of 04 2019 DAC Universal Touch Operating Instructions English...
Страница 123: ......