Dentsply Sirona
Operating Instructions DAC Universal
5 Cleaning, maintenance and validation
5.1 Check & Clean process
6681006 D3708
D3708.201.01.01.02 04.2019
93
04:00 PM
Ä
The DAC Universal pumps water for 10 seconds.
Ä
Water comes through the 6 holes in the chamber.
3.
Repeat the process if no water appears.
IMPORTANT
If the water does not flow as described above, please contact your
authorized dealer.
5.1.11
Open chamber / remove Check & Clean lid
Frequency: Weekly or every 50 cycles (whichever comes first).
ü
The DAC Universal is switched on and is in the Check & Clean
process menu.
1.
Go a step further in this Check & Clean option by pressing the
"
Navigate to next step" function key.
Ä
The lid moves up.
04:00 PM
2.
Remove the Check & Clean lid from the holder on the DAC
Universal.
5.1.12
Check the water supply for all Blue and Green lids
Frequency: Weekly or every 50 cycles (whichever comes first).
This function is used to check whether there are blockages within the lid
and their adapters.
For the following Check & Clean steps, the DAC Universal only serves
to guide the user through the Check & Clean process. Nothing else is
performed on the DAC Universal itself.
IMPORTANT
The water supply to the lid must now be checked for the Blue lid and
the Green lid as these are only fitted with internal water channels.
ü
The DAC Universal is switched on and is in the Check & Clean
process menu.
ü
All available Blue and Green lids are available and have cooled
down.