6 Errors / Display messages
6.6 Errors without a message
Dentsply Sirona
Operating Instructions DAC Universal
120
6681006 D3708
D3708.201.01.01.02 04.2019
6.6
Errors without a message
Description
Corrective action
There is water in the chamber after end
of cycle.
Drain the chamber and start a cycle.
If the error persists, clean or change the drain filter.
Check whether the drain hose is blocked or damaged.
Empty the drain tank if it is full.
If the error occurs repeatedly, please contact your authorized dealer.
Adapter is loose.
Tighten the adapter with the Allen key included.
The water tank is not automatically re-
filled.
Check whether the automatic water intake is activated in the DAC Uni-
versal menu, see "Technical settings [→ 57]".
Check whether water from the external water supply is fed to the DAC
Universal.
If the error occurs repeatedly, please contact your authorized dealer.
Loud continuous alarm signal.
Switch off the DAC Universal immediately by unplugging the power
cord.
Please contact your authorized dealer.
IMPORTANT: Do not use the device until it has been serviced by your
authorized dealer.
NOTE
Safe operation of the DAC Universal cannot be ensured unless
service intervals are observed.