Dentsply Sirona
Operating Instructions DAC Universal
4 Operation
4.2 Reprocessing with the DAC UNIVERSAL
6681006 D3708
D3708.201.01.01.02 04.2019
73
4.2.6
Reprocessing with the Green lid
The Green lid is used to clean and thermally disinfect the Green lid
instruments.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second Green lid.
WARNING
Risk of contamination!
After a reprocessing cycle is canceled, the instruments are not
reprocessed and cross-contamination can occur. Please reprocess
the instruments again.
4.2.6.1
Start cycle with the Green lid
Preparation
ü
All the presettings for the DAC Universal have been made.
ü
The DAC Universal is switched on and ready for operation.
ü
The maximum number of 6 instruments are ready for reprocessing
(for loading, see "Functions [→ 19]").
ü
The Green lid is available and cooled down.
1.
Place the Green lid for configuration on the counter or put it in the
lid holder.
2.
Attach each instrument or screw on each ultrasonic tip incl. the
torque wrench on the corresponding adapter in the Green lid.
Further information on attaching/screwing on the instruments and
replacing the adapters can be found in the separate adapter
operating instructions.
3.
Turn over the Green lid so that the instruments are facing
downwards. The green mark should be pointed towards you.
4.
Place the Green lid at the front of the DAC Universal holder and
insert the Green lid straight in the holder.
CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.
NOTE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure
there are no objects in the chamber. The DAC Universal, the lid and/
or the instruments can be damaged if a lid is not inserted correctly.