Dentsply Sirona
Operating Instructions DAC Universal
5 Cleaning, maintenance and validation
5.1 Check & Clean process
6681006 D3708
D3708.201.01.01.02 04.2019
103
04:00 PM
2.
Assemble the water tank by inserting the handle, the lid and the
float including the float guide.
04:00 PM
3.
If you do not have an automatic water connection for your DAC
Universal, fill demineralized/distilled water (water conductivity < 3.0
µS/cm) into the water tank (min. 1.0 liter, max. 2 liter).
04:00 PM
4.
Insert the water tank back into the DAC Universal.
IMPORTANT
When inserting make sure that the guide hooks on the top rear side of
the water tank are inserted correctly into the housing.
5.1.21
Preparations after the Check & Clean process
Frequency: Weekly or every 50 cycles (whichever comes first).
ü
The DAC Universal is switched on and is in the Check & Clean
process menu.
ü
NitraClean cleaning tablets are available.
CAUTION
Risk of poisoning!
Only use the NitraClean cleaning tablets stated below
1.
Go to the next Check & Clean option by pressing the "
Next Check &
Clean step" function key.