5 Cleaning, maintenance and validation
5.1 Check & Clean process
Dentsply Sirona
Operating Instructions DAC Universal
94
6681006 D3708
D3708.201.01.01.02 04.2019
ü
The Check & Clean cap is available.
1.
Go to the next Check & Clean option by pressing the "
Next Check &
Clean step" function key.
2.
Or go a step further in this Check & Clean option by pressing the
"
Navigate to next step" function key.
3.
Place the Check & Clean cap on the Blue lid.
4.
Go a step further in this Check & Clean option by pressing the
"
Navigate to next step" function key.
5.
Open the front flap of the DAC Universal.
04:00 PM
6.
Remove the water tank by carefully pulling the water tank up and
out of the DAC Universal by the handle.
7.
Place the water tank on a flat surface.
8.
Remove the lid of the water tank.
04:00 PM
9.
Remove some water from the tank with a syringe (at least 10 ml).
WARNING
Risk of clogging!
Only use demineralized or distilled water for the check, otherwise
blockages can occur, which can damage the lid and therefore prevent
the cleaning of the instruments.
04:00 PM
10. Inject the water with the syringe through both holes into the
channels of the Blue lid (as shown in the image).
Ä
Water flows through the small holes of all adapters.