41
C
ONFIGURE THE
S
ETUP
T
ABLE
The setup table consists of several sections where the attributes of the
ELITE
pro
XC
are defined.
Logger Type
Setup Table Name
Data Interval
Line Frequency
ELITE
pro
XC is the default
entry. This field cannot
be edited once you are in
the Setup Table window.
Logger type is chosen
from the
Select A Setup
Table Type
window.
Enter a name for the
setup table. 20 character
limit. This is helpful for
identifying different
setup tables.
NOTE:
The
setup table name need
not be the same as the
.
sut
file name.
The default integration
period is fifteen minutes.
The choices are 1, 3, 15,
30 seconds; 1, 2, 5, 10,
15, 20 and 30 minutes; 1
and 12 hours; 1 day
.
This
tells the logger at what
time intervals data is to
be stored in the memory.
For example, if the
integration period is set
for 30 minutes and
average watts is being
monitored, then every 30
minutes the logger
records the average
power use (watts) for
that channel over the
preceding 30 minute
interval, based on
approximately 14,400
measurements of the
monitored power draw. If
Maximum (and/or
Minimum) values are
being recorded the
highest (and/or lowest)
of those 14,400 readings
are also saved.
Auto, 60 Hz, or 50Hz.
Default is 60 Hz.
www.
.com
1.800.561.8187
Содержание ELITEpro XC
Страница 58: ...58 Typical Single Phase 2 Wire Setup Diagram www com information itm com 1 800 561 8187 ...
Страница 59: ...59 Typical Single Phase 3 Wire Setup Diagram www com information itm com 1 800 561 8187 ...
Страница 60: ...60 Typical 3 Phase 3 Wire Delta Setup Diagram www com information itm com 1 800 561 8187 ...
Страница 61: ...61 Typical 3 Phase 4 Wire Wye Setup Diagram www com information itm com 1 800 561 8187 ...