IN: 6-10 VDC, 500 mA
OUT: 6 VDC, 200 mA maximum
USB port
SENSOR LIMITATIONS
USE ONLY SHUNTED CURRENT TRANSFORMERS (CTs).
Do not use other CTs. Only use shunted CTs with a 333mV maximum output only.
Serious shock hazard and logger damage can occur if unshunted CTs are used. The
UL listing covers the use of the following DENT Instruments CTs that are UL
Recognized and have been evaluated to IEC 61010-1:
CT-RGT12-XXXX (solid core), CT-SRS-XXX (split core), CT-HSC-020-X (20A Mini), CT-
HSC-050-X (50A Mini), CT-HMC-0100-X (100A Midi), CT-HMC-0200-X (200A Midi),
CT-RXX-1310-U (R
ō
Coil), CTRXX-A4-U (R
ō
Coil), CT-CON-1000-X, CT-CON-0150EZ-X
and the CT-SRL-XXX.
The use of any other CT will invalidate the UL Listing of the
ELITE
pro
XC
.
Pulse: Use only "dry contact" non-energized pulse inputs (ELITE
pro
SP only).
Use
of energized pulse initiators can cause damage to the logger and a potential shock
hazard to the user. Wiring must meet 600V AC CAT III rating.
DC VOLTAGES
The meter is designed to measure AC voltages for installations up to 600 VAC. The
meter is also capable of measuring DC voltages up to 600 VDC with the following
caveats.
1) The internal fuse installed in the meter is not a suitable disconnect for voltages
above 80 VDC. Customers using the meter for high voltage DC systems must install
a UL listed inline 600 volt DC fuse with a rating of 1 amp or less. These fuses are
commonly available in the solar power market.
2) UL testing of the meter was conducted on AC systems only. DC results are
comparable but not tested.
3) DENT Instruments ELOG software does not currently have a configuration that is
specific to DC systems. Values reported for voltage, current, KW, KWH will be
correct however it should be noted that the meter will also report values that
pertain to AC systems (such as VARs and Power Factor) that should be ignored.
The equipment is protected throughout by double insulation (IEC 536 Class II). CAT
III 80-600 VAC 125mA 50/60 Hz
www.
.com
1.800.561.8187
Содержание ELITEpro XC
Страница 58: ...58 Typical Single Phase 2 Wire Setup Diagram www com information itm com 1 800 561 8187 ...
Страница 59: ...59 Typical Single Phase 3 Wire Setup Diagram www com information itm com 1 800 561 8187 ...
Страница 60: ...60 Typical 3 Phase 3 Wire Delta Setup Diagram www com information itm com 1 800 561 8187 ...
Страница 61: ...61 Typical 3 Phase 4 Wire Wye Setup Diagram www com information itm com 1 800 561 8187 ...