32
To measure outputs pulse:
1) To count pulses, connect the wires of your measurement device to the +/- connection points on the
ELITE
pro
SP
.
NOTE:
You can open ELOG 15, and ensure that your PC and logger time are synchronized by selecting
Logger
>
Logger Clock
>
Synchronize Time to Match PC
.
2) Using an existing setup table (or follow the steps in Section 2 of this manual to create a new one),
turn on the Digital 1 and/or Digital 2 channels.
3) Enter a descriptive identifier in the
Name
field. Choose your
Power Channel
to measure, the desired
option from the
Value
drop-down list, and enter a scaling factor in the
Scale
field.
4) Click SEND SETUP TABLE to Logger.
NOTE
: When logging is turned on or a new setup table is sent, pulse outputs start at the integration
period relative to the hour that logging starts. For example, if you have a 5 minute integration period
and you turn logging on at 1:05:01 PM (5 minutes and 1 second after the hour), logging AND pulse
output accumulation will NOT start until 1:10:00 PM.
5) Upon completion of your pulse output measurements, disconnect the wires of your device.