vPad-IV OPERATORS MANUAL
4.5.5 Graph Display
The
Graph
button as shown in
Figure 35
changes the test screen from numerical readouts to a realtime
plot of acquired data. With the graph, users can view the fluctuations and overall trends of the infusion test
(
Figure 36
).
The measured value is plotted against the elapsed time of the test. As data is acquired, the horizontal time
axis will automatically expand and the vertical axis will automatically scale so that the acquired data fits
within the graph boundaries.
Graph options depend on the test in progress, as follows:
Graph Type
Applicable Tests
Graph Description
Instantaneous
Flow
Flow Rate,
Piggyback/KVO and
PCA Pump Tests
Flow measured over short intervals ∆T ranging from 10
to 60 seconds, as set by the instantaneous flow window
of the More Settings menu (Section 4.8.7.1, page 71).
Average Flow
Flow Rate Test
Ratio of the total volume delivered to the elapsed test
time
Average &
Instant.
Flow Rate Test
A combined graph of both instantaneous and average
flow rates
Average % Error
Flow Rate Test
The error of the average flow as compared to the ml/h
rate setting of the device under test
Pressure
Occlusion Test
Pressure developed by the device under test into a
blocked IV line
Figure 36
Operation/Chapter 4
#
Page 48
Содержание vPad-IV
Страница 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Страница 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...
Страница 98: ...vPad IV OPERATORS MANUAL Appendix A Page 82 ...