vPad-IV OPERATORS MANUAL
Check Editor
A
check
step typically displays a message which requests the user to inspect the infusion device and
choose a result from the available options (PASS; FAIL; N/A; etc). Similar to a
prompt
, a
check
step
optionally provides
bold
or
small
font.
Test Editors
A test step executes an infusion device test in automatic mode as
part of the AutoSequence. Flow Rate, Piggyback, PCA and
Occlusion Tests are available. Users can add multiple tests of each
type to the AutoSequence, limited only by practical considerations
such as the total run time of the sequence. See Section 4.3.3 on page
29 for detailed instructions on configuring each type of test. Note
that manual testing is not possible in an AutoSequence.
Back Pressure Editor
A
backpr
step in an AutoSequence enables users to monitor a back pressure or vacuum that has been
externally applied to the infusion device via the output port of the Module. As the step is being executed,
users may adjust the applied pressure or vacuum until a value required by the test protocol is reached. A
back pressure step typically preceeds a Flow Rate Test where the response of the infusion device to the
applied pressure or vacuum is evaluated. The back pressure editor requires input of the target back pressure
or vacuum and the units of measure.
Pressure Monitor Editor
A
prmon
step in an AutoSequence is similar to
backpr
in that pressure in the channel is monitored
during the step. Although there is no pressure target specified for the step, the editor allows the user to
select the units of measure.
Operation/Chapter 4
#
Page 70
Содержание vPad-IV
Страница 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Страница 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...
Страница 98: ...vPad IV OPERATORS MANUAL Appendix A Page 82 ...