vPad-IV OPERATORS MANUAL
4.8 System Setup
The vPad-IV
Setup
menu (
Figure 43
), entered from the main menu of the app, enables users to access
AutoSequence programming and additional device and app settings.
4.8.1 Password Setup
Access to the
Setup
menu may be optionally protected by a password. Department managers or institutions
may require such security to prevent technicians from making changes to approved maintenance protocols.
If a password has been created, then a user must enter the correct password before the
Setup
menu will be
displayed.
To create or modify the password, press
Edit Password
on the
Setup
menu to display the "Edit
Password" dialog. An active password can be disabled by clearing the password field on dialog then pressing
Save and Exit
. As indicated on the dialog, you can long-press on the text field to erase the password.
In the event a password is entered but is later forgotten, you can enter an override password which will
grant access to the
Setup
menu. Once you gain access to the menu, the password can be erased or a new
password entered. The override password is specific to each vPad unit. Contact Datrend Customer Service
to obtain further instructions for the override password.
Figure 43
Operation/Chapter 4
#
Page 61
Содержание vPad-IV
Страница 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Страница 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...
Страница 98: ...vPad IV OPERATORS MANUAL Appendix A Page 82 ...