vPad-IV OPERATORS MANUAL
2.6 Test Monitoring
The vPad-IV app provides various display modes for monitoring a test in progress. Users can switch
between numerical readouts and a real-time graphical representation of acquired measurements. For
analyzers with multiple Modules, the "View All" display shows an overview for all enabled channels which
includes the current status and measured values for each channel.
2.7 AutoSequences
AutoSequences combine one or more automated tests of flow, volume or occlusion pressure with
user-customizable prompts and inspection checkpoints. An AutoSequence can be selected from the Test
Setup menu in the same manner as an automatic test or a manual test.
2.8 Test Records
Infusion device tests and AutoSequences create
Test Records
which are saved on the tablet. Test Records
are files which can be later reviewed, printed, or exported to a computerized maintenance management
system (CMMS). Test Records are identified by the ID or "control number" of the infusion device, which is
entered at the start of each test or AutoSequence. When channels are
linked
for testing of a multi-pump
infusion device, the results from the linked channels are combined into a single Test Record identified by
the device ID. Users may view and manage Test Records with the
vPad Record Manager
, a standard app
included with each vPad product.
2.9 Equipment Records
Each time vPad-IV completes a test and saves a Test Record, an
Equipment Record
is also saved.
Equipment Records are identified by the ID or "control number" of the infusion device, which is entered at
the start of each test or AutoSequence. Equipment Records may include the device manufacturer; model;
serial number; location; facility or owner; and so on as optionally entered by the technician at the start of a
test. In addition to the equipment information, each Equipment Record contains a reference to a
test setup
that was used to evaluate the device.
When a new test is started, input of a device ID or control number causes the vPad-IV app to search for a
corresponding Equipment Record. If an Equipment Record is found, the user is given the option of
recalling the equipment information and test setup from the Record. This saves time when starting a new
test for an infusion device which has been previously tested by vPad-IV.
Overview/Chapter 2
#
Page 10
Содержание vPad-IV
Страница 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Страница 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...
Страница 98: ...vPad IV OPERATORS MANUAL Appendix A Page 82 ...