vPad-IV OPERATORS MANUAL
1
Specifications
Equipment Performance Specifications
1.1 Instrument Specifications
1.1.1 Flow Rate Measurement
Method .......................................... Flow is calculated by measuring volume over time
Range ............................................. Measured: 0.1 ml/h to 3000 ml/h
Displayed: 0.01 ml/h to 3200 ml/h
Maximum display resolution ...... 0.001 ml/h
Accuracy ....................................... ±(1% + 0.005 ml/h) at 0.1-9.9 ml/h for volume > 100 µl;
±1% at 10-1200 ml/h for volume > 1 ml;
±2% at 1201-3000 ml/h for volume > 10 ml
1.1.2 Volume Measurement
Method .......................................... Volume is measured directly in sample sizes of 20 µl
Range ............................................. 0 to 9999 ml
Maximum display resolution ...... 0.01 ml
Accuracy ........................................ ±(1% + 1 LSD) at 0.1-9.9 ml/h for volume > 100 µL;
±(1% + 1 LSD) at 10-1200 ml/h for volume > 1 ml;
±(2% + 1 LSD) at 1201-3000 ml/h for volume > 10 ml
1.1.3 PCA Bolus or Piggyback Dose Measurement
Method .......................................... See volume measurement above
Min. detectable bolus or dose .... 0.5 ml
Bolus or dose uncertainty ........... ±10 µL for volume > 0.5 ml
PCA pump trigger output
Physical ............................ 3.5 mm monaural phono jack
Electrical .......................... SPST normally-open contact, rating 30V/300 mA
Specifications/Chapter 1
#
Page 1
Содержание vPad-IV
Страница 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Страница 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...
Страница 98: ...vPad IV OPERATORS MANUAL Appendix A Page 82 ...