vPad-IV OPERATORS MANUAL
1.1.4 Pressure Measurement
Range
General ...................................... -260 to +2600 mmHg
Back Pressure ............................ -100 to +300 mmHg operational range
Accuracy ........................................ ±(1% + 5 mmHg)
Selectable units ............................. mmHg, psi, bar, mbar, kPa, in.H
2
O, torr
Barometric compensation ........... 715 to 1095 mbar
1.1.5 Time Measurement
Elapsed Time Range .................... 0 to 240 hours
Elapsed Time Accuracy ............... ±0.5 seconds
Elapsed Time Format .................. HHH:MM:SS
PCA Lockout Time Range .......... 1 to 240 minutes
PCA Lockout Time Accuracy ..... ±0.05 minutes
1.1.6 Fluid Ports
Input ............................................... Luer lock female
Output ............................................ Hose barb, for 1/8" I.D. tubing
Material ........................................... Stainless or nickle-plated steel
1.1.7 Communication Interface Ports
Controller Module
To Tablet ................................... USB-A receptacle
PC Remote ................................ USB-B receptacle
vPad XBUS ............................... RJ12-6p6c
Controller and Accessory Modules
vPad-IV bus .............................. RJ45-8p8c
Specifications/Chapter 1
#
Page 2
Содержание vPad-IV
Страница 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Страница 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...
Страница 98: ...vPad IV OPERATORS MANUAL Appendix A Page 82 ...