Z-Trak2 3D Profiler Sensors
Z-Trak2 features reference
•
87
Display Name
Feature & Values
Description
Device
Version
& View
Line Inverter
LineInverter
Inverts the polarity of the selected line.
1.00
Beginner
Input Line Detection Level
lineDetectionLevel
Specifies the voltage threshold required to
recognize a signal transition on an input line.
1.00
Beginner
DFNC
TTL
TTL Valid detection range for 3.3/5.0 Volt (set
point 1.6 Volt).
12 Volt
input12V Valid detection range for 12 Volt (set point 6
Volt).
24 Volt
input24V Valid detection range for 24 Volt (set point 12
Volt).
Input Line Debouncing
Delay
lineDebouncingPeriod
Specifies the minimum delay (in microseconds)
to allow the input line voltage to stabilize
before determining signal transition.
1.00
Beginner
DFNC
Output Line Source
outputLineSource
Selects which internal signal to output on the
selected line.
1.00
Beginner
DFNC
Off
Off Line output is Open
Software Controlled
SoftwareControlled The Output Line Value feature changes the
state of the selected output pin.
Pulse on: Start of Frame
PulseOnStartofFrame Generate a pulse on the start of the Frame
Active event.
Pulse on: Start of
Exposure
PulseOnStartofExposure Generate a pulse on the ExposureStart event.
This option is typically used to trigger a strobe
light.
Pulse on: End of Exposure
PulseOnEndofExposure Generate a pulse on the ExposureEnd event.
This option is typically used to trigger a strobe
light.
Pulse on: Start of Readout
PulseOnStartofReadout Generate a pulse on the ReadoutStart event.
Pulse on: End of Readout
PulseOnEndofReadout Generate a pulse on the ReadoutEnd event.
Pulse on: Valid Frame
Trigger
PulseOnValidFrameTrigger Generate a pulse on the FrameTrigger event.
Pulse on: Rejected
Frame(s) Trigger
PulseOnInvalidFrameTrigger Generate a pulse on the Invalid FrameTrigger
event.
Pulse on: Start of
Acquisition
PulseOnStartofAcquisition Generate a pulse when the AcquisiontStart
event occurs.
Pulse on: End of
Acquisition
PulseOnEndofAcquisition Generate a pulse when the AcquisiontStop
event occurs.
Pulse on: End of Timer 1
PulseOnEndofTimer1 Generate a pulse on the TimerEnd 1 event.
Pulse on: End of Counter
1
PulseOnEndofCounter1 Generate a pulse on the CounterEnd 1 event.
Pulse on: Input 1 Event
PulseOnInput1 Generate a pulse on the Input signal 1 event.
Pulse on: Input 2 Event
PulseOnInput2 Generate a pulse on the Input signal 2 event.
Pulse on: Action 1
PulseOnAction1 Generate a pulse on the GigEVision Action
Command 1.
Pulse on: Action 2
PulseOnAction2 Generate a pulse on the GigEVision Action
Command 2.
Pulse on: Software
Command
PulseOnSoftwareCmd Generate a pulse on the Input of a Software
Command.
Exposure Active
ExposureActive Generate a signal that is active when the
Exposure is active.