CyberOptics Corporation
151
Gantry Components
Adjusting X-axis and Y-axis Orthogonality
This procedure is to adjust the X-axis and Y-axis orthogonality. This section shows the
procedures performed when the machine is not in the production line.
➢
To adjust the X-axis and Y-axis orthogonality
1. Turn off all electrical power to the system. For more information, refer to the section
2. Place the square granite ruler and indicators. Contact CyberOptics for the part number of the
granite ruler and indicator.
3. Make Y-axis as the reference.
4. Set a target within 5 μm and adjust the X-Axis.
Granite Ruler
Indicators
Y Indicators
screw adjustment
Target set within 5 µm