EZT-570S User Manual
57
Reverse:
When reverse action is turned on, the digital (customer event) output assigned to the
alarm will be energized in a non-alarm condition, i.e., fail-safe. When reverse action
is turned off, the output is normally open and will close in an alarm condition.
Silent:
This is used to control the method by which an alarm is reported. When silent mode
is turned off, an alarm condition will trigger the audible alarm on the chamber and the
alarm icon will flash to indicate the alarm condition. The audible alert must then be
manually silenced by pressing the reset button on the alarm monitor screen.
If silent mode is turned on, there will be no audible alert when an alarm condition is
present. In addition, if the alarm is not set to latch, the alarm icon will not be shown.
The alarm will be reported on the alarm screen, but no visual alert will be given. If
the alarm is set to latching and/or to shut down the chamber, the alarm icon will be
shown in order to provide a visual alert to the operator that user intervention is
required.
Shut Down:
If shut down is turned off, the alarm will not affect the operation of the chamber. If
shut down is turned on, the chamber will turn off all systems in an alarm condition. If
an automatic ramp/soak program is operating, the program will also go into hold.
Digital Output Assignment
If the EZT-570S is equipped with the customer event option, the alarm can be assigned to one of the
event outputs. When the alarm activates, the output will work in conjunction with the alarm and turn
on when the alarm is active and off when the alarm is cleared.
The event output must be configured to be used as an alarm output. If the alarm output
selected is not configured to be an alarm output, it will not function according to the alarm
settings. See Section 8.6.4 Digital Outputs, for information on how to configure the
operational mode of the customer event outputs.
5.4.1
Set Points and Alarm Message
The alarm set points and alarm message are accessed by pressing the “Setpoints” button at the
bottom of the Alarm screen.
Figure 5-5 Alarm Set Points and Message