EZT-570S User Manual
111
8.6.3.1 Digital
Input
Wiring
Locate the digital input connector on your chamber. Figure 8-16 is a representation of the typical
digital input connection provided on most chambers; however, as changes and/or improvements are
made, your specific connection style may vary.
It is recommended that all wiring be of the shielded type. Only connect the drain wire of the shielded
wiring at your device, or the chamber, not both. Connecting the shield to ground at both ends of the
cable can induce unwanted “noise” into the input wiring. Improper and/or poor wiring practices can
cause interference on the digital inputs which may cause erratic or uncontrolled operation.
Figure 8-16 Digital Input Wiring
The digital inputs will work with either transistor or potential free (relay) contacts. The input
connection provides both 24Vdc power and common terminals for use with the digital inputs. Power
can also be provided by a separate source. In order for a separate source to work properly, the
common terminal of the external power supply must be connected to the common terminal on the
input connector.
Do not connect the +24Vdc terminal on the chamber to the output of another power
source or in series with an external source. Doing so can cause irreparable damage
to the EZT-570S and may result in a hazardous shock or fire.
Do not exceed the digital input voltage range or damage to the input may occur. The
+24Vdc supply is of limited power and for sourcing the inputs only. Do not use this
to power other devices or short the +24Vdc and common terminals together. Damage
to the power supply may occur.
Digital Input Specifications
Voltage Input
Input Range
20.4 to 28.8Vdc
Input Impedance
3.4K ohm
Rated Input Current
7mA/point
Response Time
1 second maximum