EZT-570S User Manual
18
1.5 Notifications
The notification window is a feature that can be accessed by pressing the date/time field in the icon
bar. This window provides a snapshot of current activity. The notification window can be closed by
pressing the date/time field again, or is automatically closed if the Menu, Home, Alarm or left or right
arrow icon is pressed.
Figure 1-5 Notifications
All available notifications can be viewed by touching the screen and dragging a finger up or down on
the list to scroll through all items like all other lists in the EZT-570S. The notifications include the
following:
•
Loop PV and SP for each loop.
•
Monitor PV for each monitor point (if option installed on chamber).
•
Automatic program status including name of running program.
•
Defrost operational status (if chamber equipped with defrost).
•
Active alarm status including the most recent alarm.
•
Data logging status including active file name and the length and logging rate set.
•
Security status including the current user.
•
Audit trail active/inactive status.
•
Web server active/inactive status.
•
VNC server active/inactive status.
•
EZT-570S IP address.
•
Available SD storage memory remaining.
The notification items do not update while the list is shown. The information shown in the list
is a snapshot of the data at the time the window was shown. Therefore, the PV or SP shown
for a loop may not be the actual values for the loop if the notification window is left open for
extended periods of time. The notification window is not available under offline settings.