EZT-570S User Manual
110
Previous Step:
When selected, the input will advance the ramp/soak program to the next
step. The program must first be placed in hold prior to the input turning
on in order for the step change to take place.
Next Step:
When selected, the input will advance the ramp/soak program to the
previous step. The program must first be placed in hold prior to the input
turning on in order for the step change to take place.
*
Chamber Outputs Off:
When selected, the input will cause all chamber control outputs to turn
off. The outputs will stay off as long as the input is on. When the input
turns off, all outputs will return to normal operation.
*
All Outputs Off:
When selected, the input will cause all chamber control and event
outputs to turn off. This includes all optional digital outputs configured as
customer events. The outputs will stay off as long as the input is on.
When the input turns off, all outputs will return to normal operation.
*
Event Outputs Off:
When selected, the input will cause all optional digital outputs configured
as customer events to turn off. The outputs will stay off as long as the
input is on. When the input turns off, all event outputs will return to
normal operation.
*
Event Output Control:
When selected, the input becomes a “switch” which can be used to
control optional digital outputs. When a digital output is configured to be
controlled by the digital input, it will turn on when the input is on and off
when the input is off.
*
For the output control modes, the “on” state for the input is defined by the transition mode.
If “High” is selected, the input is on when voltage is applied to the input. If “Low” is selected,
the input is on when voltage is removed from the input.