Select ‘Specify an IP address’ and enter in the desired IP address, subnet mask and default gateway
if required. For the 4.3” display version of the EZT-570S, you can move the keyboard around on the
screen to position it as necessary to access the different fields. Once complete, close the ‘Ethernet
Drivers’ window by pressing the ‘OK’ button located at the top right of the window. Next, close the
Connection window by pressing the ‘X’ button at the top right of the window.
Access the taskbar and hide the keyboard by pressing the keyboard button at the bottom right of the
taskbar and select ‘Hide Input Panel’. DO NOT cycle power to the EZT-570S at this time. Wait
approximately 2 minutes before removing power. This provides time to allow the Windows CE
operating system to save the new network settings to the registry so that it uses the settings on the
next reboot.
If power is cycled too soon, the network settings will revert back to the previous settings on the next
power up. Once the 2 minute time period has elapsed since updating the network settings, cycle
power to the chamber in order to restart the system and return to normal operation with the new
network settings.
9.3.3
Using the Web Server
The EZT-570S’s web server allows a user to remotely monitor chamber operations anywhere via a
PC’s standard web browser. When enabled, the web server provides a detailed status page of the
EZT-570S’s current operating conditions. The web server can be enabled and disabled by simply
touching the “Www.” button.
You must use the assigned IP address to access the EZT-570S. To obtain the correct IP address, go
to the Communications screen selection under the device settings “Comm” menu. Write down the IP
address so you will have it to enter into your web browser. You can also view the IP address by
pressing the date/time field in the menu bar area to show the notifications window. The IP address is
shown at the bottom of the notifications list.
Figure 9-9 Web Server Settings
Contact your IT department prior to enabling the web server of the EZT-570S. Company
policy may prohibit the use of web servers for security reasons. We are not responsible for
the use of, nor makes any claims as to the security of the web server interface over your
network. The use of the EZT-570S web server is the responsibility of the end user.