EZT-570S User Manual
117
8.7.1
Refrigeration Monitor Option
If the chamber is equipped with the refrigeration monitor option, the monitor inputs will be used to
display the suction and discharge temperatures and pressures of the refrigeration system
compressors. The pressure sensor inputs are in absolute pressure (PSIA), not gauge pressure
(PSIG). In order to display the refrigeration system pressures in gauge pressure, the offsets for the
pressure monitor inputs will be set to a default value of 14.7 (standard atmospheric pressure). This
converts the absolute pressure readings to gauge pressure.
The EZT-570S uses 9300 controllers to monitor the refrigeration system temperatures and
pressures. One 9300 is used for each suction and discharge line. Input 1 of the 9300 is
temperature and input 2 is pressure. The monitor input offset can be used to correct for slight
errors in readings; however, actual calibration of the inputs must be performed at the 9300.