COM Server Commands
76
SCPI.CALCulate(
Ch
).SELected.MARKer(
Mk
).FUNCtion.DOMain.
STARt
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel
Ch
(if the marker search range coupling is set to
OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.
DOMain.COUPle
property), the active trace of channel
Ch
(if
otherwise),
Ch
Description
The start value of the marker search range.
Range
From the stimulus current start value to the stimulus current stop
value.
Out of Range
No limitation
Preset Value
3e5
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Syntax
Dim Value As Double
Value =
app.SCPI.CALCulate(Ch).SELected.MARKer(1).FUNCtion.
DOMain.STARt
app.SCPI.CALCulate(Ch).SELected.MARKer(1).FUNCtion.
DOMain.STARt = 1e6
Equivalent
Softkeys
Markers > Marker Search > Search Start