COM Server Commands
228
SCPI.SENSe(
Ch
).CORRection.EXTension.PORT(
Pt
).FREQuency(
Ls
)
Object Type
Property (read/write)
Data Type
Double
Target
Port
Pt
of channel
Ch
,
Ch
Pt
port number 1–2 (see
Description
The value of the frequency at the point number
Ls
for calculation of the
loss for the port extension function,
Ls
Range
from 2E4 to 4.8E9
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1e9
Unit
Hz (Hertz)
Syntax
Dim Value As Double
Value =
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).
FREQuency(Ls)
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).
FREQuency(Ls) = 100Е6
Equivalent
Softkeys
Calibration > Port Extensions > Loss > Freq1 | Freq2
Table 9. Ls: point number
Data Type
Long
Description
Point number for setting the losses in the functions extension of port
Range
1 or 2
Out of Range
An error occurs. Error code: 222