COM Server Commands
248
SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel
Ch
,
Ch
Description
The ON/OFF state of the continuous frequency offset adjust feature.
Allowable values
True
continuous offset adjust feature is ON
False
continuous offset adjust feature is OFF
Preset Value
False
Syntax
Dim Status As Boolean
Status =
app.SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe
app.SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe =
True
Equivalent
Softkeys
Stimulus > Freq.Offset > Offset Adjust > Contin.Adjust
SCPI.SENSe(Ch).OFFSet.ADJust.EXECute
Object Type
Method
Target
Channel
Ch
,
Ch
channel number 1–9 (see Table 1)
Description
Executes the offset adjust when the frequency offset feature is ON
Syntax
app.SCPI.SENSe(Ch).OFFset.ADJust.EXECute
Equivalent
Softkeys
Stimulus > Freq.Offset > Offset Adjust > Adjust Immediate