COM Server Commands
315
SCPI.TRIGger.SEQuence.EXTernal.POSition
Object Type
Property (read/write)
Data Type
String
Target
External trigger
Description
Selects the position of the external trigger. The Analyzer waits for
external trigger:
Before sampling, when the frequency of the stimulus port have been
set. The frequency change of the stimulus port begins after sampling.
Before the frequency setup and subsequent measurement. The
frequency change of the stimulus port begins when the external trigger
arrives.
Depending on the command
SCPI.TRIGger.SEQuence.POINt
the
external trigger wait occurs before each point or before the first point
of the full sweep cycle.
Range
"BSAM"
Before sampling
"BSET"
Before frequency setup
Out of Range
An error occurs.
Preset Value
"BSAM"
Syntax
Dim Param As String
Param = app.SCPI.TRIGger.SEQuence.EXTernal.POSition
app.SCPI.TRIGger.SEQuence.EXTernal.POSition =
"BSAM"
Related
Commands
SCPI.TRIGger.SEQuence.SOURce
Equivalent
Softkeys
Stimulus > Trigger > External Trigger > Position > Before Sampling |
Before Setup