COM Server Commands
208
SCPI.SENSe(
Ch
).CORRection.COLLect.CKIT.STAN(
Std
).Z0
Object Type
Property (read/write)
Data Type
Double
Target
Standard
Std
of the calibration kit specified for channel
Ch
,
Ch
Std
standard number (see
Description
The offset Z0 value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50 or 75, depending on the selected calibration kit.
Unit
Ω (Ohm)
Syntax
Dim Value As Double
Value =
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN
(Std).Z0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN
(Std).Z0 = 50
Equivalent
Softkeys
Calibration > Edit Cal Kit > Offset Z0
SCPI.SENSe(
Ch
).CORRection.COLLect.CLEar
Object Type
Method
Target
Channel
Ch
,
Ch
Description
Clears the measurement values of the calibration standards.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.CLEar
Equivalent
Softkeys
Calibration > Calibrate > Response (Open) | Response (Short) | Response
(Thru) | One Path 2–Port Cal | Full 1–Port Cal > Cancel