![Copper Mountain Technologies PLANAR TR1300/1 Скачать руководство пользователя страница 132](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189132.webp)
COM Server Commands
132
SCPI.DISPlay.WINDow(
Ch
).ANNotation.MARKer.ALIGn.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Channel
Ch
,
Ch
Description
The alignment mode of the marker display position of each trace, when
the only active trace display feature is turned OFF
(
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.S
TATe
property is set to
False
).
Range
"VERTical"
Vertical alignment
"HORizontal"
Horizontal alignment
"NONE"
No alignment
Notes
The short format of the parameter is indicated by upper case letters.
There is no distinction between upper and lower case letters when the
property is written. When the property is read out, the short format is
indicated by upper case letters.
Out of Range
The value is ignored.
Preset Value
"NONE"
Syntax
Dim Param As String
Param =
app.SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.
ALIGn.TYPE
app.SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.
ALIGn.TYPE = "VERT"
Equivalent
Softkeys
Markers > Properties > Align > Vertical | Horizontal | OFF