COM Server Commands
192
SCPI.SENSe(
Ch
).CORRection.COLLect.ACQuire.THRU
Object Type
Property (write only)
Data Type
Variant (Long array)
Target
Channel
Ch
,
Ch
Description
Measures the calibration data of the thru standard between the source
port and the receiver port.
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same
port numbers.
Out of Range
If an incorrect port number is specified, an error occurs. If the same port
numbers are specified, an error occurs.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU=
Array(2, 1)
Notes
The property writing starts the measurement for the channel
independently of the trigger initiation and trigger source settings. The
function of the property writing waits for the completion of the
measurement.
Equivalent
Softkeys
Calibration > Calibrate > Response (Thru) > Thru
Calibration > Calibrate > One Path 2–Port Cal > Thru