COM Server Commands
152
SCPI.INITiate.CONTinuous
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the continuous trigger initiation mode. If the
continuous trigger initiation mode is set to OFF, the channel turns to
the hold state.
Allowable
Values
True
Continuous trigger initiation mode ON
False
Continuous trigger initiation mode OFF
Preset Value
True
Syntax
Dim Status As Boolean
Status = app.SCPI.INITiate.CONTinuous
app.SCPI.INITiate.CONTinuous = False
Notes
The sweep start in continuous trigger initiation mode depends on the
trigger source. If the trigger is set to internal, the sweeps will go
immediately one after another. If the trigger is set otherwise, the sweep
will start when the trigger signal is received.
Equivalent
Softkeys
Stimulus > Trigger > Continuous
Stimulus > Trigger > Hold