COM Server Commands
34
SCPI.CALCulate(
Ch
).SELected.CONVersion.FUNCtion
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel
Ch
,
Ch
Description
The S–parameter conversion function type.
Range
"IMPedance"
Reflection or Transmission equivalent impedance
according to the trace measurement parameter (S11
or S21)
"ADMittance"
Reflection or Transmission equivalent admittance
according to the trace measurement parameter (S11
or S21)
"INVersion"
Inverse S–parameter
"CONJugation"
S–parameter conjugate
Notes
The short format of the parameter is indicated by upper case letters.
There is no distinction between upper and lower case letters when the
property is written. When the property is read out, the short format is
indicated by upper case letters.
Out of Range
An error occurs.
Preset Value
"IMP"
Syntax
Dim Param As String
Param =
app.SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion
app.SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion
= "INV"
Equivalent
Softkeys
Analysis > Conversion > Function > Impedance Z | AdmittanceY | Inverse
1/S | Conjugation