COM Server Commands
143
SCPI.DISPlay.WINDow(
Ch
).TRACe(
Tr
).Y.SCALe.RLEVel
Object Type
Property (read/write)
Data Type
Double
Target
Trace
Tr
of channel
Ch
,
Ch
Tr
trace number 1–8 (see
Description
The value of the reference line (response value on the reference line).
For the rectangular format only.
Range
from –1E–18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0 (except for SWR: 1)
Unit
dB (decibel) | ° (degree) | s (second)
Syntax
Dim Value As Double
Value =
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.
RLEVel
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.
RLEVel = 10
Equivalent
Softkeys
Scale > Ref Value