![Copper Mountain Technologies PLANAR TR1300/1 Скачать руководство пользователя страница 218](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189218.webp)
COM Server Commands
218
SCPI.SENSe(
Ch
).CORRection.COLLect.ECAL.SOLT1
Object Type
Property (write only)
Data Type
Long
Target
Channel
Ch
,
Ch
Description
Executes Full 1-port calibration of the specified port of specified
channel (
Ch
) using the AutoCal module.
Allowable
Values
Port number = 1
Syntax
Dim Port As Long
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT1 =
Port
Equivalent
Softkeys
Calibration > AutoCal > Full 1-Port Calibration