COM Server Commands
114
SCPI.CALCulate(
Ch
).SELected.TRANsform.TIME.STIMulus
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel
Ch
,
Ch
Description
The selection of the stimulus type for the time domain transformation
function: impulse or step.
Range
"IMPulse" Impulse
"STEP"
Step
Notes
The short format of the parameter is indicated by upper case letters.
There is no distinction between upper and lower case letters when the
property is written. When the property is read out, the short format is
indicated by upper case letters.
Out of Range
The value is ignored.
Preset Value
"IMP"
Syntax
Dim Param As String
Param =
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.
STIMulus
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.
STIMulus = "STEP"
Equivalent
Softkeys
Analysis > Time Domain > Response Type > Bandpass | Lowpass Step |
Lowpass Impulse