![Copper Mountain Technologies PLANAR TR1300/1 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189069.webp)
COM Server Commands
69
SCPI.CALCulate(
Ch
).SELected.MARKer(
Mk
).ACTivate
Object Type
Method
Target
Marker
Mk
of the active trace of channel
Ch
,
Ch
Mk
marker number 1–15, or reference marker number 16 (see
Description
Sets the active marker. If a marker is OFF this function will turn it ON.
Turning ON a marker with the number from 1 to 15 will turn ON all the
markers of smaller numbers. Turning ON the reference marker with
number 16 does not turn ON the markers with the numbers from 1 to
15, but switches these markers to the relative measurement mode.
Syntax
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate
Equivalent
Softkeys
Markers > Active Marker > Marker n
Markers > Reference Marker
Table 4. Mk: Marker Number
Data Type
Long
Description
Marker number. Numbers from 1 to 15 are for regular markers, number
16 is for the reference marker.
Range
from 1 to 16
Out of Range
An error occurs..
Notes
If the marker number is not specified, it is taken as equal to 1.