![Copper Mountain Technologies PLANAR TR1300/1 Скачать руководство пользователя страница 251](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189251.webp)
COM Server Commands
251
SCPI.SENSe(Ch).OFFSet.ADJust.Value
Object Type
Property (read/write)
Data Type
Double
Target
Channel
Ch
,
Ch
channel number 1–9 (see Table 1)
Description
The frequency offset adjust value of the offset adjust feature
Range
from –5E5 to 5E5
Out of range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
Hz
Syntax
Dim Value As Double
Value = app.SCPI.SENSe(Ch).OFFSet.ADJust.VALue
app.SCPI.SENSe(Ch).OFFSet.ADJust.VALue = 1000
Equivalent
Softkeys