COM Server Commands
187
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1
Description
Selects the port and sets the full 1–port calibration type, when the
written calibration coefficients are made effective by the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE
method.
Type
Long (write only)
Range
Port number is 1 or 2. (see Table 2)
Out of Range
An error occurs. Error code: 222.
Target
Channel
Ch
,
Ch
Syntax
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.
SOLT1 = Port
Equivalent
Softkeys
None