CI-154 User’s Manual
Page iii
ADDENDUM
Portable Particle Counters
Security System Enhancements
ACCESS TO CLEAR DATA MENU BUTTON
Firmware version 3.20 has been created to enable more control over the availability
of the CLEAR DATA button depending on the security level of the user that is logged
in.
There are four levels of security available on the portable particle counters.
Analysts with Level 1 security can operate a portable particle counter to select ID’s
and acquire data. Level 1 users cannot print reports or clear data. Previous
firmware versions permit Level 2 or 3 users to print a report but they can also clear
data.
Starting with firmware version 3.20 it is possible to remove the ability for Level 2
and 3 users to clear data.
A special RS-232 command now exists, which a technician can send one time to the
counter. After the command has been received, the [CLEAR DATA] button will be
greyed out for Level 2 and 3 users and they will no longer be able to clear data.
This characteristic will be retained permanently by the counter unless it is cancelled
by another special RS-232 command.
Please contact Climet Instruments Co. or your Authorized Climet service provider
for further information.
BACKDOOR SUPERVISOR PASSWORD
Climet Instruments Co. and authorized Climet service providers are able to disclose
a secret supervisor password to enable a supervisor to gain access to their counter
if the password set by the supervisor has been forgotten. Starting with version
3.00, a program was developed that enables a temporary password to be
generated that is only good for one particular counter until its calibration date
changes. Starting with version 3.02, the backdoor password was changed so the old
universal one no longer works.
If the supervisor password is forgotten, please feel free to contact Climet or an
authorized Climet service provider. Please provide the firmware version number,
serial number and calibration date exactly as it is shown on the splash screen.
Please also provide the name, title and contact information for the person at your
company who is authorized to receive the temporary password.
Содержание CI-154
Страница 2: ...CI 154 User s Manual Page ii ...
Страница 4: ...CI 154 User s Manual Page iv ...
Страница 9: ...CI 154 User s Manual Page ix FRONT VIEW OF CI 154 ...
Страница 10: ...CI 154 User s Manual Page x REAR PANEL OF CI 154 ...
Страница 16: ...CI 154 User s Manual Page 6 ...
Страница 18: ...CI 154 User s Manual Page 8 ...
Страница 80: ...CI 154 User s Manual Page 70 ...
Страница 84: ...CI 154 User s Manual Page 74 ...
Страница 92: ...CI 154 User s Manual Page 82 ...
Страница 96: ...CI 154 User s Manual Page 86 ...
Страница 98: ...CI 154 User s Manual Page 88 ...