CI-154 User’s Manual
Page 18
Validation reports for ISO 14644-1, GMP, or FS-209E can be generated by the CI-
154. ISO reports conform to ISO 2015 starting at firmware version 3.30. When
performing validation monitoring, each sampling location in an area should be
identified by selecting the appropriate ID for that location. While ISO 14644-1 allows
a single location (for example, when sampling a clean hood), FS-209E requires that
a minimum of two locations be sampled. Failure to have at least two location IDs
within the stored data will cause the area to fail.
The CI-154 will generate ISO 14644-1, GMP, or FS-209E reports using all sample
data that is in memory at the time the report is requested. For a given report, the
samples used to make up that report will all be consecutive with the same program
name. A change in program name generates a separate report. The number of
reports generated depends on the type of report and the settings, but there will
always be at least one plus the number of program name changes within the stored
data (The same program name appearing again later counts as a change).
Before beginning the sample process, clear stored data if desired so that previously
collected data with the same program name is not included with the current
calculations, and reports from programs of no interest are not generated. At each
location, select the appropriate ID by pressing [
ID
] from the main data screen, then
highlight the appropriate ID from the list using the [
PAGE UP
] and [
PAGE DOWN
] keys to
search a screen at a time, and the [
UP SELECT
] and [
DOWN SELECT
] key to highlight the
ID. Press [
EXIT
] to accept the new ID. Once the ID is selected, take the specified
number of samples for that location, then move to the next location and repeat the
process. Once all the data has been collected, the report can be printed. It is
possible to automatically select the next ID on the list by enabling [
AUTO ADVANCE
].
To print the report, press [
REPORT
] from the main data screen, then press
[
ISO/209E/GMP
] from the bottom of the report control screen. You can view the report
in the preview screen that appears, using the [
PAGE UP
] and [
PAGE DOWN
] keys to
navigate through multiple screens if the report is long. Press [
] to print the
report. Press [
EXIT
] if you do not want to print the report.
GMP differs from ISO 14644-1 and FS-209E in a number of respects, one of them
being that GMP uses the 95% UCL for less than 10 locations and allows one location
with outlying data—very low counts, for example—to be suppressed from the
calculations, to reduce the standard deviation.
To suppress one location from the report, press [
EXCLUDE LOCATION
], which takes you
to the ID menu. Highlight the ID to exclude it, then press [
EXIT
]. To restore the ID,
press [
EXCLUDE LOCATION
], then press the [
NO EXCLUDE
] key. There must be at least
three measurement points remaining, and the justification for excluding the
outlier
must be documented.
Printing alarm violation reports
The CI-154 can print out a report of all alarm violations for the data still in memory.
To print the report, press [
REPORT
] from the main data screen, then press [
ALARM
DATA
] in the following screen. To view the alarm data on screen, press the [
PAGE UP
]
and [
PAGE DOWN
] keys. To print the report, press [
].
2.2.11
How to view stored data on-screen
End of sample
data can be viewed from the screen without printing out the data.
From the main data screen press [
REPORT
], then press [
ALL DATA
] from the report
Содержание CI-154
Страница 2: ...CI 154 User s Manual Page ii ...
Страница 4: ...CI 154 User s Manual Page iv ...
Страница 9: ...CI 154 User s Manual Page ix FRONT VIEW OF CI 154 ...
Страница 10: ...CI 154 User s Manual Page x REAR PANEL OF CI 154 ...
Страница 16: ...CI 154 User s Manual Page 6 ...
Страница 18: ...CI 154 User s Manual Page 8 ...
Страница 80: ...CI 154 User s Manual Page 70 ...
Страница 84: ...CI 154 User s Manual Page 74 ...
Страница 92: ...CI 154 User s Manual Page 82 ...
Страница 96: ...CI 154 User s Manual Page 86 ...
Страница 98: ...CI 154 User s Manual Page 88 ...