CI-154 User’s Manual
Page 3
When cleaning agents are sprayed near the instrument, be sure that
the instrument is off, and that the inlet is capped.
1.3.6
Storage
If the CI-154 is to be stored for more than one month without use, the battery
should be disconnected to prevent damage to the battery from a deep discharge.
Please refer to the Service and Maintenance section of this manual for instructions
on long term storage.
1.3.7
Packing for shipment
The packing container used for shipping the CI-154 has been tested according to the
standards of the National Safe Transit Association. For this reason, it is
recommended that the original packing container be saved for reuse if it has no
obvious damage. The packing container consists of an outer double-walled
cardboard box, two foam inserts, and an internal cardboard box used to hold
accessories. This internal box also serves the important function of tightening the
foam inserts against the CI-154. If for any reason the internal box cannot be used,
insert additional packing material to fill the empty space and tighten the foam
inserts against the CI-154.
NOTE:
When returning the CI-154 for servicing, keep the
accessories in a safe place at your facility. No
accessories are needed during calibration or repair.
If the original shipping carton was discarded, select a cardboard box large enough
that at least 3 inches of heavy bubble wrap can be placed on
all
sides of the
instrument. Any extensions (handle, feet) should be especially well protected,
because the impact is amplified if the carton is dropped on a part that extends out
from the rest of the chassis. Avoid the use of loose packing materials such as
peanuts
, because they shift and leave one side of the instrument unprotected.
Shipping companies do not accept responsibility for damage when the instrument is
poorly packed.
1.4
Overview of CI-154 features
The CI-154 (Climet Instruments 154) is a laser diode based aerosol particle counter
that monitors particles in four size ranges.: 0.3, 0.5, 1 and 5
m. It is fully self-
contained, operating on battery power or AC power for sampling convenience. It has
a stainless steel enclosure that makes it suitable for applications requiring a sterile
wipe-down.
The CI-154 has a sample flow rate of 1 cubic foot per minute (1 CFM) or 28.3 liters
per minute (28.3 LPM), requiring 35.3 minutes to take a full 1 cubic meter sample.
The built-in battery can provide 5.5–6.5 hours of continuous sampling on a single
charge. The CI-154 interfaces with the user through a color LCD touch screen. It
comes with a thermal printer, and uses patented optical and flow systems with
automatic flow control. The sample air is filtered before being exhausted through the
rear panel. When used for special applications where the air sample must be
exhausted out of the area, a fitting can be attached to the exhaust port to allow
tubing to be connected to the exhaust.
Содержание CI-154
Страница 2: ...CI 154 User s Manual Page ii ...
Страница 4: ...CI 154 User s Manual Page iv ...
Страница 9: ...CI 154 User s Manual Page ix FRONT VIEW OF CI 154 ...
Страница 10: ...CI 154 User s Manual Page x REAR PANEL OF CI 154 ...
Страница 16: ...CI 154 User s Manual Page 6 ...
Страница 18: ...CI 154 User s Manual Page 8 ...
Страница 80: ...CI 154 User s Manual Page 70 ...
Страница 84: ...CI 154 User s Manual Page 74 ...
Страница 92: ...CI 154 User s Manual Page 82 ...
Страница 96: ...CI 154 User s Manual Page 86 ...
Страница 98: ...CI 154 User s Manual Page 88 ...