Appendix
Appendix 6 Mixer Measurement
399
Fig. 6.25 [Vector mixer calibration] Dialog Box
5) [Help] dialog box for saving of mixer characteristics
The vector network analyzer has the function to save characteristic data of the calibration mixer. After calibration,
the mixer characteristic calibration file will be saved to the specified path. If the same calibration mixer is used in
vector mixer measurement calibration, the saved S2P file can be recalled directly.
Fig. 6.26 [Save characteristics] Dialog Box
Browse:
find the path to save the characteristic data of the calibration mixer. Use the default or custom name.
Next:
click [Next] to complete system calibration after saving the characteristics of the calibration mixer.
6.6 Embedded LO Mixer Measurement
A fixed LO is embedded in the tested mixer. The characteristics of the embedded LO can be applied in vector
mixer, scalar mixer measurement, IMDX, IMXSpectrum, GCX and NFX measurement.
It is challenging to measure such devices for the following two reasons.
1) The reference mixer with the same LO frequency is required in vector mixer measurement. The independent
internal source or external source must be used as LO of the reference mixer. At the same time, LO is controlled by
the instrument. The measurement velocity is higher if the internal source 2 is used.
2) All the embedded LO measurement must be adjusted with the analyzer receiver to the exact frequency to
measure the mixer output. Thus the LO frequency must be accurate.
1) How to measure the embedded LO?
Enter the nominal frequency of the embedded LO in the [Mixer setting] dialog box, as the start point of
measurement.
Before measurement sweep of the DUT, start background sweep to determine the frequency of the embedded LO
obtain the configurable accuracy. Background sweep:
1) Broadband sweep: initially measure the embedded LO frequency, and add one frequency range near the selected
data point. The input signal of the DUT must be selectively adjusted to CW. Receiver B is used to sweep the
selected frequency range near the preset output frequency. Obtain the desired difference between the output signal
and the found frequency, and apply the obtained result in adjustment.
2) Accurate sweep: Receive B is used to measure the selected data point. Measure the ratio of the phase and time,
and calculate the exact frequency offset, until reaching the allowable residual value or maximum degree of
repetition.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...