11 Technical Indicators and Measurement Methods
11.2 Measurement Methods
323
21) After all calibrations, click [OK]. Then click [Response] > [Avg] > [Average Factor] and enter 8 to enable the
averaging function.
22) The maximum value of the displayed curve within each frequency range is the dynamic range of Port 3.
Fig. 11.9 Block Diagram I of Dynamic Range
Calibration Test of System
Fig. 11.10 Block Diagram II of Dynamic Range
Calibration Test of System
Test of dynamic range of Port 4:
23) Click [Meas] > [S43].
24) Repeat Step (18) to (22) to test the dynamic range of Port 4.
11.2.10 Test of Transmission Tracking, Reflection Tracking, Effective Source Matching,
Effective Direction, etc.
11.2.10.1 Items of Test of Transmission Tracking, Reflection Tracking, Effective Source Matching, Effective
Direction and Effective Load Matching
Test the indicators of transmission tracking, reflection tracking, effective source matching, effective direction and
effective load matching after the full double-port calibration with AV3672 series vector network analyzers.
11.2.10.2 Instrument and Accessories for Test of Transmission Tracking, Reflection Tracking, Effective
Source Matching, Effective Direction and Effective Load Matching
1)
3.5mm coaxial Calib calibration kit AV31121
2)
Soft test cable
Agilent85131F
11.2.10.3 Steps of Test of Transmission Tracking, Reflection Tracking, Effective Source Matching, Effective
Direction and Effective Load Matching
1) As shown in Fig. 11.11, connect the system and preheat is for 60min at least.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...