Appendix
Appendix 5 Amplifier Gain Compression Measurement
379
➢
Receiver Attenuator
: refer to the attenuator specified as the receiver to be connected to the DUT output
port.
➢
Source Leveling Mode
: select the specified internal amplitude stabilization mode.
Power sweep:
➢
Power Points
: refer to the number of power points required for two-dimensional sweep measurement.
The power point number may be limited by the frequency point number. This item is not included in
intelligent sweep, as the compression level is found based on sufficient power points in the intelligent sweep
mode.
➢
Start and stop power
X/Y sweep
: for the return, X/Y and maximum gain compression, the reference gain and compression point are
measured within the power range. Therefore, the range should be wide enough to include the reference gain
and compression point.
For example, if the return level is 10dB, the power range must be more than 10dB.
Otherwise, the closest reference gain and compression point will be used to determine the compression value in
amplifier gain compression, which will lead to inaccurate results. In the linear gain compression, the reference gain
is measured at the linear power level, and the start and stop power levels are used to find the compression point.
Intelligent sweep
: the power range is set to be larger than the compression point searched in amplifier gain
compression.
The reference gain found in compression is applied to the linear power level. The return and X-axis
value are determined according to the compression method. The repeated sweep to find the compression point
should be reduced.
Prompt
Low-band high-power mode
If DUT needs higher input power to realize compression below 3.2GHz, the high power mode should be used.
Click [Path configuration] > [Low-band high-power mode].
➢
Power step:
calculated according to the current start, stop and point number settings. It cannot be changed
directly.
Path configuration:
Open the [Path configuration] dialog box.
5.3.3.3 Compression Setting
Menu path:
[Response] > [Measure] > [Gain Compression]
. Click
[Compression]
to set compression. After
setting, click
[OK]
to close the dialog box.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...