6 Measurement Optimization
6.2 Improvement of Reflection Accuracy of Low-loss 2-port Device
184
Menu path:
[Analysis] > [Time Domain] > [Time Domain Transformation...]
.
Click the
[Transform]
check box in the
[Transform]
dialog box to activate the transformation mode. Click the
required transformation mode: low-pass impulse, low-pass step and band pass. Click
[Frequency: low pass]
to
select the low pass mode. Click the drop-down list in the
gate type
box to select the gate type: band pass and trap.
Set the gate threshold by setting the
start
value,
stop
value, center value or
span
. Click
[Transform]
in the
trace
menu.
6.2 Improvement of Reflection Accuracy of Low-loss 2-port Device
In order to realize accurate reflection measurement in one-port calibration, connect the unmeasured port. This
method is particularly suitable for low-loss two-way devices, such as the filter passband and cable. The direction,
source matching and frequency response error can be corrected by means of one-port calibration. The load
matching error cannot be calibrated. Therefore, the load matching error must be minimized.
Connect the unmeasured port in the following methods:
➢
Connect the unmeasured port of the DUT to one high-performance terminal load (the calibration kit is also
allowed). The measurement accuracy of this method is close to that of 2-port calibration.
➢
Connect the unmeasured port of the DUT directly to the analyzer. Insert one 10dB precision attenuator
between the device output and analyzer. The effective load matching of the analyzer can be improved by once
based on the measured value of the attenuator, i.e. 20dB.
Prompt
Matched load
No consideration should be given to load matching in measurement of the device of high reverse isolation, such as
the amplifier.
Fig. 6.1 shows the instability of measurement of the DUT with or without 10dB precision attenuator at the end.
Fig. 6.1 Comparison of Instability
Refer to the filter reflection.
Refer to the instability of measurement without 10dB precision attenuator.
Refer to the instability of measurement with 10dB precision attenuator.
Fig. 6.2 shows the accurate reflection measurement of the 2-port device. The following calculation shows the
improvement of loading matching with one high-quality 10dB attenuator.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...