11 Technical Indicators and Measurement Methods
11.1 Technical Indicators
309
11 Technical Indicators and Measurement Methods
This chapter describes the technical indicators and measurement methods of AV3672 series vector network
analyzers.
●
Technical indicators
·······························································································
●
Measurement methods
·····························································································
11.1 Technical Indicators
This section lists the details of performance indicators and technical parameters of AV3672 series vector network
analyzers. By reading this section, the user will have an accurate understanding on main performance indicators of
the product. Normal working conditions of the instrument include the ambient temperature of 0°C to +40°C and 1
standard atmospheric pressure.
Main technical parameters of AV3672 series vector network analyzers are given below. The temperature of test
environment should be 23±3°C, and the ambient temperature change must not exceed ±1°C after calibration. The
pressure should be the atmospheric pressure.
Attention
Different models and options:
AV3672A/B/C/D/E are respectively used to test the frequency of 13.5GHz/26.5GHz/43.5GHz/50GHz
/67GHz
.
AV3672A/B/C/D/E of standard configuration can test the indicators of Port 1 and 2. The four-port type
AV3672A/B/C/D/E can test the indicators of Port 1, 2, 3 and 4.
11.1.1 Frequency Range
Frequency range: AV3672A: 10MHz to 13.5GHz
AV3672B: 10MHz to 26.5GHz
AV3672C: 10MHz to 43.5GHz
AV3672D: 10MHz to 50GHz
AV3672E:10MHz ~ 67GHz
11.1.2 Frequency Accuracy
Frequency accuracy: ±1×10
-7
(23
℃
±3
℃
)
11.1.3 Frequency Resolution
Frequency resolution: 1Hz
11.1.4 Port Output Harmonic Suppression
Table 11.1-1 Indicators of AV3672A/B Port Output Harmonic Suppression
(Power setting: -5dBm, 2/3 harmonics)
Frequency range
10MHz
≤
f
≤
4GHz
4GHz
<
f
≤
26.5GHz
Output harmonic suppression of
Port 1 and 3
-
48dBc
-
57dBc
Output harmonic suppression of
Port 2 and 4
-
13dBc
-
18dBc
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...